AmplifiersAnalyzersCable & Antenna AnalyzersCablesCalibration KitsCalibratorsCounters / TimersDetectorsEmulatorsError DetectorsLogic AnalyzersMetersModules / Plug InsMultimetersNetwork AnalyzersNoise GeneratorsNoise SourcesOpticalOptical Spectrum AnalyzersOscilloscopesOtherPattern GeneratorsPower MetersPower SensorsPower SuppliesProbesRF / MicrowaveService Monitors / Test SetsSignal AnalyzersSignal GeneratorsSpectrum AnalyzersThermal ImagersVXI
We buy, sell and rent general purpose, optical and RF test and measurement equipment. We pay top dollar for your unwanted, surplus or defective gear. Let us make you an offer on your new or used Agilent, Tektronix, Anritsu, Fluke, Rohde & Schwarz and IFR instruments.
Item Description
You are bidding on a Brand New in the box Tektronix 80E09B 60 GHz Electrical Dual Channel Sampling Module with Remote Samplers. It is a high-frequency, low-noise signal acquisition and jitter characterization solution compatible with the Tek DSA8200/DSA8300 Sampling Oscilloscopes. It will arrive with all the standard accessories and a fresh calibration performed by Tektronix, to include full test data! The 80E09B has a current list price from the manufacturer over $60,000 USD!
The DSA8300 Series Sampling Oscilloscope, when configured with one or more electrical sampling modules, provides complete test solutions for multi-channel, high-bandwidth electrical applications. This highly configurable solution is well suited for debugging, characterizing and analyzing components, modules and systems with signaling rates at 10, 40 and even 100 Gb/s. The 80E00 family of electrical sampling modules provide a wide variety of capabilities, allowing the user to configure a test solution specifically adapted to their application.
The 80E09B and 80E07B are dual-channel modules with remote samplers, capable of 450 μVRMSnoise at 60 GHz sampling bandwidth, and 300 μV at 30 GHz sampling bandwidth. Each small form factor remote sampler is attached to a 2-meter cable to minimize the effects of cables, probes, and fixtures, allow close location of the sampler to the DUT, and enable best signal fidelity. User-selectable bandwidth settings (60/40/30 GHz on 80E09B and 30/20 GHz on 80E07B) offer optimal noise/bandwidth trade-off.
Key Features & Specifications:
-Up to 70 GHz bandwidth and 5 ps measured rise time (10-90%)
-Lowest noise for analysis: 450 μVRMS at 60 GHz, 300 μVRMS at 30 GHz
-15 ps reflected true differential fully integrated TDR rise time (12 ps incident) and feature resolution below 1 mm (TDR modules)
-Efficient, accurate, easy to use, and cost-effective S-parameters up to 50 GHz
-Remote samplers 1 enable location of sampler near DUT and ensure best signal fidelity
80E09B Specs:
Rise time - 5.8 ps (calculated from 0.35 bandwidth rise time product )
Dynamic range - 1.0 Vp-p
Offset range - ±1.1 V
Maximum operating voltage - ±1.1 V
Maximum Nondestruct Voltage, DC + ACp-p - 2.0 V
Vertical number of digitized bits - 16 bits full scale
Vertical sensitivity range - 10 mV to 1.0 v full scale
DC vertical voltage accuracy, single point, within ±2 °C of compensated temperature - ±2 mV
± 0.007 * (Assigned Offset)
± 0.02 * (Vertical Value - Assigned Offset)
Step response aberrations, typical
±1% or less over the zone 10 ns to 20 ps before step transition
+6%, –10% or less for the first 400 ps following step transition
+0%, –4% or less over the zone 400 ps to 3 ns following step transition
+1%, –2% or less over the zone 3 ns to 100 ns following step transition
±1% after 100 ns following step transition
RMS noise, maximum (warranted, typical)
60 GHz: ≤600 μV (450 μV)
40 GHz: ≤480 μV (330 μV)
30 GHz: ≤410 μV (300 μV)
Value in parenthesis is typical value to which the instrument will normally perform
This unit is in excellent condition, has been fully tested and is guaranteed functional. It comes with a fresh calibration and a 14 day right of return should it not perform to your expectations. Please review pics and text carefully before bidding to verify accuracy of description. If you don't see it in the pictures it is not included.
Insurance is included in UPS shipping charges. Item shipped internationally on buyer's UPS, FEDEX or DHL accounts only.
We will accept payment for this item by company check, cashier’s check and wire transfer only.
If you have zero feedback, contact us prior to purchase so we know you are a serious bidder. Otherwise, we may cancel your bid before the auction closes. It should be obvious that you should only bid if you are serious and do intend to complete the transaction. Any stated warranty will be voided if the unit is opened or if any calibration or lockout stickers have been tampered with or removed.
Please note that shipping and insurance costs are non-refundable.
Our feedback is very important to us and we strive for 100% positive on every transaction. If you are unsatisfied with any aspect of your transaction, please contact us before leaving negative or neutral feedback so that we can address and remedy the issue immediately. Additionally, we are always happy to leave positive feedback for our trading partners after a successful transaction.
Make sure that you check out our other auctions for more great deals on test and measurement equipment.
Thank you and good luck bidding.
Need your unit calibrated by an ISO 17025 accredited lab prior to shipment? Not a problem. We work with numerous calibration and repair companies that can provide your required service at competitive prices.